PurposeTo develop a time-efficient pulse sequence that acquires multiple diffusion-weighted images with distinct diffusion times in a single shot by using multiple stimulated echoes (mSTE) with variable flip angles (VFA). MethodsThe proposed diffusion-weighted mSTE with VFA (DW-mSTE-VFA) sequence begins with two 90 degrees RF pulses that straddle a diffusion gradient lobe (G(D)) to excite and restore one half of the magnetization into the longitudinal axis. The restored longitudinal magnetization was successively re-excited by a series of RF pulses with VFA, each followed by another G(D), to generate a set of stimulated echoes. Each of the multiple stimulated echoes was acquired with an EPI echo train. As such, the train of multiple stimulated echoes produced a set of diffusion-weighted images with varying diffusion times in a single shot. This technique was experimentally demonstrated on a diffusion phantom, a fruit, and healthy human brain and prostate at 3 T. ResultsIn the phantom experiment, the mean ADC measured at different diffusion times using DW-mSTE-VFA were highly consistent (r = 0.999) with those from a commercial spin-echo diffusion-weighted EPI sequence. In the fruit and brain experiments, DW-mSTE-VFA exhibited similar diffusion-time dependence to a standard diffusion-weighted stimulated echo sequence. The ADC showed significant time dependence in the human brain (p = 0.003 in both white matter and gray matter) and prostate tissues (p = 0.003 in both peripheral zone and central gland). ConclusionDW-mSTE-VFA offers a time-efficient tool for investigating the diffusion-time dependency in diffusion MRI studies.
Keyword:
DWI
stimulated echo
time-dependent diffusion MRI
variable flip angle

期刊

Magnetic Resonance in Medicine 封面图
Magnetic Resonance in Medicine
IF:
3
论文数:
1.2W
被引数:
3.1W

机构

University of Illinois System 封面图
University of Illinois System
学者数:
6.9W
论文数: 6.2W
被引数: 644
引用论文

引用论文

Diffusion-prepared 3D gradient spin-echo sequence for improved oscillating gradient diffusion MRI
err2020-07-08
err22
PREAI
errWu, Dan; Liu, Dapeng; Hsu, Yi-Cheng; Li, Haotian; Sun, Yi; Qin, Qin; Zhang, Yi
err分享
err收藏
Quantification of cell size using temporal diffusion spectroscopy
err2015-04-04
err89
errOAAI
errJiang, Xiaoyu; Li, Hua; Xie, Jingping; Zhao, Ping; Gore, John C.; Xu, Junzhong
err分享
err收藏
Surface-to-Volume Ratio Mapping of Tumor Microstructure Using Oscillating Gradient Diffusion Weighted Imaging
err2015-07-24
err60
errOAAI
errReynaud, Olivier; Winters, Kerryanne Veronica; Dung Minh Hoang; Wadghiri, Youssef Zaim; Novikov, Dmitry S.; Kim, Sungheon Gene
err分享
err收藏
err分享
err收藏
High-Spatial-Resolution Diffusion MRI in Parkinson Disease: Lateral Asymmetry of the Substantia Nigra
errRADIOLOGY
IF15.2
err2019-04-01
err51
errOAAI
errZhong, Zheng; Merkitch, Douglas; Karaman, M. Muge; Zhang, Jiaxuan; Sui, Yi; Goldman, Jennifer G.; Zhou, Xiaohong Joe
err分享
err收藏
Diagnostic value of diffusion-weighted STEAM-MRI in ischemic stroke弥散加权STEAM MRI对缺血性脑卒中的诊断价值
err2021-06-01
err3
PREAI
errMueller, Sebastian Johannes; Khadhraoui, Eya; Van Kube, Julia My; Langer, Philip; Riedel, Christian Heiner; Voit, Dirk; Ernst, Marielle; Frahm, Jens
err分享
err收藏
学者 查看更多内容