Multicontact Localization Framework for Flexible Robots Using a Data-Driven Approach

delete2023-12-01
delete0
delete
OA
AI
X
Xuan Thao Ha *
A
Aditya Sridhar
M
Mouloud Ourak
G
Gianni Borghesan
A
Arianna Menciassi
E
Emmanuel Vander Poorten
DOI:10.1109/JSEN.2023.3326614delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
Minimally invasive surgery (MIS) is increasingly employed in several medical disciplines. During minimal invasive cardiovascular approaches, flexible instruments, such as catheters or endoscopes, are navigated along a tortuous path to reach a deeply seated target site. Navigation through the vasculature is not without risk, as the vessel can be fragile, potentially with plaque or calcification that could get dislodged if excessive stresses are caused during navigation. Knowledge of the interaction forces between these flexible tools and the vasculature could help steer the instruments and avoid exerting excessive forces on the surrounding tissues. Integrating force sensing into the tip of the instruments has been proposed in the art. However, tip force sensing is blind to forces that arise along the catheter body's length. This article proposes a data-driven framework to estimate the locations of multiple contact forces along the catheter's length. The proposed approach consists of a contact state detection and, subsequently, a contact localization method. Both methods make use of knowledge of the shape of the catheter's body. This shape is measured here by a multicore fiber Bragg grating (FBG) fiber. The proposed method also works for steerable catheters, as it allows accounting for the effect of the actuation of the catheter itself. A state-of-the-art contact localization method using multicore FBG fiber is implemented and serves as a baseline to compare the newly proposed method with. The proposed data-driven approach requires a minimum amount of training data, which can be collected by simply actuating the robotic catheter in free space over its workspace. The method is validated with a commercial MitraClip steerable guide catheter. Dynamic experiments show that the proposed contact state detection method can detect contact in 0.44 s. The average contact localization errors of the proposed method and state-of-the-art method are 6.1 mm (1.6%) and 18.0 mm (4.9%), respectively.
Keyword:
Contact localization
force sensing
optical fiber
shape sensing
steerable catheters/needles

期刊

IEEE Sensors Journal 封面图
IEEE Sensors Journal
IF:
4.5
论文数:
2.1W
被引数:
7.3W

机构

K
KU Leuven
学者数:
5.7W
论文数: 5.2W
被引数: 8.1W
引用论文

引用论文

Spiral FBG sensors-based contact detection for confocal laser endomicroscopy
err2020-12-01
err16
PREAI
errGao, Anzhu; Liu, Ning; Zhang, Haojie; Wu, Zicong; Yang, Guang-Zhong
err分享
err收藏
Three dimensional force estimation for steerable catheters through bi-point tracking
err2018-08-01
err20
errOAAI
errBack, Junghwan; Lindenroth, Lukas; Rhode, Kawal; Liu, Hongbin
err分享
err收藏
Image-Based Optical Miniaturized Three-Axis Force Sensor for Cardiac Catheterization
err2016-11-01
err56
errOAAI
errNoh, Yohan; Liu, Hongbin; Sareh, Sina; Chathuranga, Damith Suresh; Wurdemann, Helge; Rhode, Kawal; Althoefer, Kaspar
err分享
err收藏
err分享
err收藏
Triaxial Catheter-Tip Force Sensor for MRI-Guided Cardiac Procedures
err2013-02-01
err108
PREAI
errPolygerinos, Panagiotis; Seneviratne, Lakmal D.; Razavi, Reza; Schaeffter, Tobias; Althoefer, Kaspar
err分享
err收藏
学者 查看更多内容